Test Entry - Detail

Anti-Insulin Antibody


General
  Clinical Use: Insulin autoantibodies may be useful in the investigation of hypoglycaemia.
If working up a new diabetic patient, suggest measure anti-GAD Ab and anti-IA2 Ab instead.
  Availability: Samples are batched and tested fortnightly.
  Aliases/Synonyms: Insulin Antibody,   Insulin Autoantibody,  
  Code: IAB
  Handling Instructions (to laboratory): Send sample chilled.
  Hyperlink:
  Reference Interval:
< 0.4 kU/L


Collection Requirements
  Container: Serum Sep. Tube-SST (GOLD),  
  Sample Type: See container
  Minimum Collection Volume: 1mL
  Collection Instructions: Requests should be reviewed by Duty Biochemist prior to collection.


Processing Requirements
  Alternate Containers:
  Processing Instructions: Centrifuge, separate and freeze sample.
  Minimum Assay Volume: 0.1mL
  Stability:
  Transport Instructions (to testing laboratory): Send separated sample frozen on dry ice.
Requests should be reviewed by Duty Biochemist prior to transport.


Testing Locations
  Performed at:
Section Department Site Contact Phone
- Laboratory SSWPS-Royal Prince Alfred Hospital (02) 9515 8279


Last Updated : 23-11-2023 13:45